EPrints@IIT Delhi >
Faculty Research Publicatons >
Physics >
Please use this identifier to cite or link to this item:
http://hdl.handle.net/2074/699
|
Full metadata record
| DC Field | Value | Language |
| contributor.author | Yashwant Kumar | - |
| contributor.author | Vinay Kumar | - |
| contributor.author | Jain, Kamlesh K | - |
| contributor.author | Kashyap, Subhash C | - |
| date.accessioned | 2005-08-05T03:59:13Z | - |
| date.available | 2005-08-05T03:59:13Z | - |
| date.issued | 1999 | - |
| identifier.citation | Sensors and Actuators B, 55(2-3), 217–221 | en |
| identifier.uri | http://eprint.iitd.ac.in/dspace/handle/2074/699 | - |
| description.abstract | This paper describes the performance of an alternative capacitive pressure gauge using lead iron tungstate as a capacitive element over a pressure range of 418 MPa. A commercial strain gauge pressure transducer is used to get the consistency in the
calibration equation of the pressure transducer using the piston gauge pressure standard when compared with the calibration
equation of the pressure transducer obtained using the capacitive pressure gauge. A close agreement better than 1=10-3 (over all accuracy of the pressure transducer is -5=10-3) demonstrate for the first time the use of the capacitive pressure gauge as a reliable pressure standard where the required accuracy in the pressure measured is below 1=10-3. | en |
| format.extent | 196323 bytes | - |
| format.mimetype | application/pdf | - |
| language.iso | en | en |
| subject | Transducer | en |
| subject | Pressure measurements | en |
| subject | Pressure standards | en |
| subject | Hydraulic pressures | en |
| title | A capacitive pressure gauge as a reliable transfer pressure standard | en |
| type | Article | en |
| Appears in Collections: | Physics
|
Files in This Item:
| File |
Description |
Size | Format |
| kumarcap1999.pdf | | 191Kb | Adobe PDF | View/Open |
|
Show simple item record
All items in DSpace are protected by copyright, with all rights reserved.
|